SAFETY SERIES No.115

INTERNATIONAL BASIC SAFETY STANDARDS FOR PROTECTION AGAINST INOIZING RADIATION AND FOR THE SAFETY OF RADIATION SOURCES

INTERIM EDITION

Jointly sponsored by:
Food and Agriculture Organization of the United Nations
Interbational Atomic Energy Agency
International Labour Organization
Nuclear Energy Agency of the
Organization for Economic Co-operation and Development
Pan American Health Organization
World Health Organization

INTERNATINAL ATOMIC ENERGY
VIENNA,1996



Schedule 3

GUIDANCE LEVELS OF
DOSE, DOSE RATE AND ACTIVITY
FOR MEDICAL EXPOSURE


GUIDANCE LEVELS FOR DIAGNOSTIC RADIOLOGICAL PROCEDURES

TABLE 3-1 GUIDANCE LEVELS OF DOSE FOR DIAGNOATIC RADIOGRAPHY FOR A TYPICAL ADULT PATIENT

TABLE 3-2 DOSE GUIADANCE LEVELS FOR COMPUTED TOMOGRAPHY FOR A TYPICAL ADULT PATIENT

TABLE 3-3 DOSE GUIADANCE LEVELS FOR A TYPICAL ADULT PATIENT

TABLE 3-4 DOSE RATE GUIADANCE LEVELS FOR FLUOROSCOPY FOR A TYPICAL ADULT PATIENT

GUIADANCE LEVELS FOR DIAGNOSTIC PROCEDURES IN NUCLEAR MEDICINE

TABLE 3-5 GUIADANCE LEVELS OF ACTIVITY FOR PROCEDURES IN NUCLEAR MEDICINE FOR A TYPICAL ADULT PATIENT

GUIADANCE LEVELS OF ACTIVITY FOR DISCHARGE FROM HOSPITAL

TABLE 3-6 GUIADANCE LEVEL FOR MAXIMUM ACTIVITY FOR PATIENTS IN THERAPY ON DISCHARGE FROM HOSPITAL


TABLE 3-1 GUIDANCE LEVELS OF DOSE FOR DIAGNOATIC
RADIOGRAPHY FOR A TYPICAL ADULT PATIENT
           
Examination
Entrance surface dose
per radioraph*
( mGy )
Lumbar spine AP 10
LAT 30
LSJ 40
Abdomen,intravenous urography    
and cholecystography
AP 10
Pelvis AP 10
Hip joint AP 10
Chest PA 0.4
LAT 1.5
Thoracic spine AP 7
LAT 20
Dental Periapical 7
AP 5
Slull PA 5
LAT 3

Notes: PA:posterior-anterior;LAT:lateral projection, LSJ:lumbo-sacral-joint projection;AP:anterior-posterior projection.

*In air with backscatter.These values are for conventinal film-screen combination in the relative speed of 200.For high speed film-screen combinations(400-600),the values should be reduced by a factor of 2 to 3.

p279

----- go top -------


TABLE 3-2 DOSE GUIADANCE LEVELS FOR COMPUTED TOMOGRAPHY FOR A TYPICAL ADULT PATIENT

Examination       
Maltiple scan average dose*
( mGy )
Head            50
Lumbar spine 35
Abdomen 25
*Derived from measurement on the axis of rotation in water equivalent phantoms, 15cm in length abd 16cm(head) and 30cm(lumber spine and abdomen) in diameter.

----- go top -------


TABLE 3-3 DOSE GUIADANCE LEVELS FOR MAMMOGRAPHY FOR A TYPICAL ADULT PATIENT

    Average glandilar dose per cranio-caudal projection*    
1 mGy (without grid)
3 mGy (with grid)
*Dtermined in a 4.5 cm compressed breast consisting of 50% glandular and 50% adipose tissue, for film-screen systems and dedicated Mo-target Mo-filter mammography units.

----- go top -------


TABLE 3-4 DOSE RATE GUIADANCE LEVELS FOR FLUOROSCOPY FOR A TYPICAL ADULT PATIENT

Mode of operation       
Entrance surface dose rate*
( mGy/min )
Normal            25
High level** 100
* In air with backscatter
** For flouroscopes that have an optional 'high level' operational mode, such as those frequently used in interventional radiology.

p280

----- go top -------