SAFETY SERIES No.115

INTERNATIONAL BASIC SAFETY STANDARDS FOR PROTECTION AGAINST INOIZING RADIATION AND FOR THE SAFETY OF RADIATION SOURCES

INTERIM EDITION

Jointly sponsored by:
Food and Agriculture Organization of the United Nations
Interbational Atomic Energy Agency
International Labour Organization
Nuclear Energy Agency of the
Organization for Economic Co-operation and Development
Pan American Health Organization
World Health Organization

INTERNATINAL ATOMIC ENERGY
VIENNA,1996



Schedule ‚R

GUIDANCE LEVELS OF
DOSE, DOSE RATE AND ACTIVITY
FOR MEDICAL EXPOSURE


GUIDANCE LEVELS FOR DIAGNOSTIC RADIOLOGICAL PROCEDURES

TABLE ‚R|‚P GUIDANCE LEVELS OF DOSE FOR DIAGNOATIC RADIOGRAPHY FOR A TYPICAL ADULT PATIENT

TABLE ‚R|‚Q DOSE GUIADANCE LEVELS FOR COMPUTED TOMOGRAPHY FOR A TYPICAL ADULT PATIENT

TABLE ‚R|‚R DOSE GUIADANCE LEVELS FOR A TYPICAL ADULT PATIENT

TABLE ‚R|‚S DOSE RATE GUIADANCE LEVELS FOR FLUOROSCOPY FOR A TYPICAL ADULT PATIENT

GUIADANCE LEVELS FOR DIAGNOSTIC PROCEDURES IN NUCLEAR MEDICINE

TABLE ‚R|‚T GUIADANCE LEVELS OF ACTIVITY FOR PROCEDURES IN NUCLEAR MEDICINE FOR A TYPICAL ADULT PATIENT

GUIADANCE LEVELS OF ACTIVITY FOR DISCHARGE FROM HOSPITAL

TABLE ‚R|‚U GUIADANCE LEVEL FOR MAXIMUM ACTIVITY FOR PATIENTS IN THERAPY ON DISCHARGE FROM HOSPITAL


TABLE ‚R|‚P GUIDANCE LEVELS OF DOSE FOR DIAGNOATIC
RADIOGRAPHY FOR A TYPICAL ADULT PATIENT
@@@@@@@@@@@
Examination
Entrance surface dose
per radioraph*
( mGy )
Lumbar spine AP 10
LAT 30
LSJ 40
Abdomen,intravenous urography@@@@
and cholecystography
AP 10
Pelvis AP 10
Hip joint AP 10
Chest PA 0.4
LAT 1.5
Thoracic spine AP 7
LAT 20
Dental Periapical 7
AP 5
Slull PA 5
LAT 3

Notes:@PA:posterior-anterior;LAT:lateral projection, LSJ:lumbo-sacral-joint projection;AP:anterior-posterior projection.

*In air with backscatter.These values are for conventinal film-screen combination in the relative speed of 200.For high speed film-screen combinations(400-600),the values should be reduced by a factor of 2 to 3.

p279

----- go top -------


TABLE ‚R|‚Q DOSE GUIADANCE LEVELS FOR COMPUTED TOMOGRAPHY FOR A TYPICAL ADULT PATIENT

Examination@@@@@@@
Maltiple scan average dose*
( mGy )
Head@@@@@@@@@@@ 50
Lumbar spine 35
Abdomen 25
*Derived from measurement on the axis of rotation in water equivalent phantoms, 15cm in length abd 16cm(head) and 30cm(lumber spine and abdomen) in diameter.

----- go top -------


TABLE ‚R|‚R DOSE GUIADANCE LEVELS FOR MAMMOGRAPHY FOR A TYPICAL ADULT PATIENT

@@@@Average glandilar dose per cranio-caudal projection*@@@@
1 mGy (without grid)
3 mGy (with grid)
*Dtermined in a 4.5 cm compressed breast consisting of 50% glandular and 50% adipose tissue, for film-screen systems and dedicated Mo-target Mo-filter mammography units.

----- go top -------


TABLE ‚R|‚S DOSE RATE GUIADANCE LEVELS FOR FLUOROSCOPY FOR A TYPICAL ADULT PATIENT

Mode of operation@@@@@@@
Entrance surface dose rate*
( mGy/min )
Normal@@@@@@@@@@@ 25
High level** 100
*@In air with backscatter
** For flouroscopes that have an optional 'high level' operational mode, such as those frequently used in interventional radiology.